4. Connect the blood line with the expansion chamber to the
dialyzer.
5. Prime and de-air the blood lines and the dialyzer as described
in chapter 4 "Hemodialysis - Double Needle Treatment" in
part 1.
5.a.
Tilt the expansion chamber holder backwards until it
clicks into position. By doing so the venous expansion
chamber will be
fi
lled to a suitable level during
priming.
5.b.
Check that the venous expansion chamber has been
fi
lled to be level with the groove of the chamber. Make
sure to close the clamp at the top of the expansion
chamber.
5:4
AK 96
®
Operator's Manual - Single Needle Treatment
HCEN12239 Revision 12.2010
Program version 3.xx