Diascan NOT POSSIBLE during
UF STEP/INTERVAL profiling
Appears:
During an attempt to activate the Diascan function when
UF pro
fi
ling in steps or intervals is active, or when an
attempt to activate UF pro
fi
ling in steps or intervals
is made and Diascan measuring has previously been
activated.
Machine actions:
The Diascan function is not available in combination with
UF pro
fi
ling in steps or intervals. The Diascan function
will automatically be de-activated.
Measures:
-
Note that the Diascan function is no longer active.
-
The attention alarm automatically disappears after
30 seconds.
Mute time:
Permanently muted buzzer.
Buzzer sound:
Soft buzzer.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:41
Attention alarms in alphabetical order
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...