2. Select alternative in the
ASSISTED PRIMING SETUP MENU
using the
Keypad
.
3.
Press the
Select key
and the menu will change.
If the dialysis
fl
uid preparation is complete
("F.Ch" disappears
from the Time Display and the bypass path on the Flow Diagram
lights up) the dialysis
fl
uid tubes can be connected to the dialyzer.
See "Dialysis Fluid Tubes - Attach", page 4:35 in part 1 for
instructions on how to attach the tubes. Especially note that if
UFD (dialysis
fl
uid
fi
lter) is installed, the dialysis
fl
uid tubes
cannot be moved until the bypass path of the Flow Diagram
lights up green.
If the dialysis
fl
uid preparation is not complete
, the process
can be started with the dialysis
fl
uid tubes connected to the safety
couplings.
4. Press the
fl
ashing
Blood pump button
to start the blood pump
and the procedure. If the blood pump segment has been
automatically threaded, check for possible kinks. Note that
the dialyzer should be
fi
lled from bottom to top to ensure
proper de-airation.
The procedure starts with the
fi
ll up phase.
ACTIVE PHASE:FILLUP
will be displayed.
5. Fill up the level in the venous drip chamber (while the
blood pump is running) using the level adjustment knob
(anticlockwise direction). Adjust the level so it is well above
the air detector head.
6. The
Air detector button
will start to
fl
ash when the
venous drip chamber has been
fi
lled. This is to request
the operator to press it and by doing so, activate the air
detector alarm function. It is not necessary to press the
fl
ashing button at this moment, the air detector activation
may be moved forward until the extracorporeal circuit is
completely de-aired. Additionally, if the air detector is not
manually activated during priming, it will be automatically
activated either when the priming volume is achieved, when
CONNECT PATIENT
is selected or when blood is detected.
The activated air detector alarm function is indicated by an
unlit button.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Double Needle Treatment
4:47
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...