IMPOSSIBLE TIME OR UF SETTING
Adjust TIME or UFV
Appears:
When a too high UF volume has been set in combination
with the set treatment time.
Machine actions:
Buzzer is activated 2 min. after attention alarm is
generated.
Measures:
Adjust Time,
SET TIME
, or UF volume,
SET UF
, until
the calculated UF rate is lower than the high limit for
UF rate.
Mute time:
2 min.
Buzzer sound:
Soft buzzer
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:77
Attention alarms in alphabetical order