3. Select
INTERVAL
below
SET AUTO MODE
using the
Keypad
to activate automatic
measuring.
4. Check that the desired
time intervals between the
measurement checks, displayed
below
SET INTERVAL
,
correspond with the displayed
ones, or change the time intervals
using the
Keypad
. The time
intervals can be set between 5–60
minutes.
5. If desired, set high and low alarm
limits for systolic, diastolic and
mean blood pressure as well
as pulse rate by
fi
rst pressing
the
Back key
to go back to the
BPM MENU
. In this menu select
ALARM
using the
Keypad
to
reach the
BPM ALARM MENU
.
5.a.
First check or set
ALARM
to
ON
using
the
Keypad
.
5.b.
Then select
SET LIMITS
and the
menu will change.
11:10
AK 96
®
Operator's Manual - BPM
HCEN12239 Revision 12.2010
Program version 3.xx