Heat Disinfection Program - Performing
Note
•
The program can be interrupted at any time when the program
is running; press and hold the
Select key
for 3 seconds. The
machine will then interrupt the procedure and
fi
nish off with
a drain sequence.
•
The pick-up tubes must be rinsed/disinfected separately. See
"Pick-up Tubes” on page 8:41 for instructions.
Note
Check before starting the procedure
That the dialysis
fl
uid tubes are connected to the safety couplings
and that the latches of the BiCart holder are closed.
That the concentrate connectors are placed in the corresponding
stand-by ports of the machine (blue and red), and that the yellow
disinfectant connector is placed in the parking port (marked P).
Procedure
Use the
Keypad
next to the Information Display to select options and
to navigate within and between the menus. For details concerning the
keys included in the
Key pad
, see "Keypad" in chapter 3. part 1
1. Press the
Rinse/Disinfection button
brie
fl
y.
2. The
RINSE/DISINFECTION MENU
will be displayed on
the Information Display.
3. Select
HEAT
.
4. Check that
HEAT
is displayed as the selected program.
If
HEAT
is displayed, continue to point 5.
If not, do point 4.a and 4.b before continuing to point 5.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Hygiene and Maintenance
8:11
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...