FCH IS WAITING
Close or check BiCart holder
Appears:
During function check if a chemical disinfection program
has previously been performed and the latches of the
BiCart holder are open.
Machine actions:
Function check is interrupted.
Measures:
Close the latches of the BiCart holder. It is not possible
to connect the BiCart cartridge yet.
Mute time:
2 min.
Buzzer sound:
Soft buzzer
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:51
Attention alarms in alphabetical order
Summary of Contents for AK 96