Procedure for manual activation
1. Press the
Fluid path button
, select
DIA FLUID FLOW
using
the
Keypad
.
2. Press the
Select key
for 3 seconds to activate concentrate
stand-by mode.
3. The
CONCENTRATE STANDBY MODE
menu will now be
displayed on the Information Display.
Resume fluid preparation
To resume the dialysis
fl
uid preparation (Concentrate stand-by
mode deactivated) press the
Select key
for 3 seconds when the
CONCENTRATE STANDBY MODE
is displayed. The dialysis
fl
uid
will be ready within approximately 2 minutes.
If dialysis
fl
uid preparation has not been resumed during priming, it
will automatically be resumed when blood is detected in the priming
detector after treatment start. Dialysis
fl
uid preparation will also
automatically be resumed when concentrate stand-by mode has been
active for more than one hour.