3. Select
CHEMICAL
.
4. Select the desired chemical disinfectant program
.
5. Start the program by pressing the
Select key
for 3seconds,
Rinse/Disinfection button
lights up.
DISINFECTION IN PROGRESS:
will be displayed and will
remain throughout the program.
The
Rinse/Disinfection button
will continue to be lit throughout
the program, indicating that the program is ongoing.
6. Connect the yellow disinfectant connector (see point 6 in
"Fluid Part Component Details", starting on page 2:19) to
the pick-up tube and place it in the disinfectant container,
alternatively move the disinfectant connector from the
parking port to the disinfectant port.
7. If the disinfectant connector is still connected to the parking
port when the
fi
lling phase of the program starts, the
following attention alarm will be generated:
Note especially that the disinfectant connector must remain in the
disinfectant container/disinfectant port and not be moved to the
parking port before the program is complete. This is due that the
chemical disinfectant is pushed back to the disinfectant container
to empty the connector/intake tube at the end of the program.
8. When the program is complete the
Priming button
starts to
fl
ash and
DISINFECTION COMPLETED:
is displayed.
9. Switch the machine off by pressing the
On/Off button
for
3 seconds.
8:28
AK 96
®
Operator's Manual - Hygiene and Maintenance
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...