5.c.
Set the proper alarm limits
suitable for the patient
using the
Keypad
. For
further information, see
starting on page 13:26 in
part 1.
Step back to the
AUTO BPM MENU
using
the
Back key
.
The
fi
rst measurement check in the sequence will be
performed when the time for the
fi
rst interval has ended,
e.g. if the interval is set to 20 minutes the
fi
rst check
will be 20 minutes after activation. The current pressure
in the blood pressure measurement cuff is displayed in
P CUFF
on the
BPM MENU
.
If the measuring process needs to be interrupted, when
the
BPM MENU
is open, press the
Select key
to stop the
procedure.
AUTO BPM
will be shown at the activity
fi
eld of the
Treatment Overview Menu when automatic BPM
measuring has been activated. During the ongoing blood
pressure measurement check, this will be replaced by a
fl
ashing
BPM
label.
Note especially that it is possible to start a single
measurement check at any time between the automatic
measurement checks if necessary. When the Treatment
Overview Menu is displayed; press the
BPM button
and
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - BPM
11:11
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...