If UF volume or treatment time is changed, after UF pro
fi
ling has
been activated, the pro
fi
ling graph will automatically change as well.
The UF rate change will if possible be kept and the graph will shift
in parallel to the original graph.
Examples of UF rate linear graphs where time or UF volume has been
changed are shown in the pictures below (changed graphs in gray,
original graph in black). Note that if time is changed and the stop value
of the new graph reaches the minimum UF rate, the slope of the graph
will also change.
Figure 7:5
Change of UF rate linear graph when UF volume has
been changed
Figure 7:6
Change of UF rate linear graph when time has been
changed
Note
•
For any UF rate graph, always check the UF pro
fi
ling settings
if UF volume or treatment time is changed after UF pro
fi
ling
activation.
Note
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Profiling
7:5