General
In order to optimize the dialysis treatment the concentrations of sodium
and bicarbonate in the dialysis
fl
uid and the ultra
fi
ltration rate can
automatically be changed following predetermined graphs.
Profiling of Na
+
and HCO
3-
dialysis fluid concentration
For sodium and/or bicarbonate concentrations, linear increasing or
decreasing graphs can be set. The graph represents a constant change
in the sodium and/or bicarbonate concentrations in the dialysis
fl
uid
throughout the treatment. The change is determined by the start and
stop values set for each parameter. Based on these values, the machine
will perform a smooth and constant adjustment of the concentrations
during the treatment time when pro
fi
ling has been activated. See the
following
fi
gure.
Figure 7:1
Linear graphs for sodium and bicarbonate pro
fi
ling
7:2
AK 96
®
Operator's Manual - Profiling
HCEN12239 Revision 12.2010