Heat Disinfection Program - Automatic Switch Off
The machine can be preset to switch off automatically after a complete
heat disinfection program, i.e. when the
Priming button
is
fl
ashing, by
an authorized technician.
It can also be switched off during the heat disinfection program, by
manually pressing the
On/Off button
while the program is running.
This will make all of the operator panel lights go out, except for the
Rinse/Disinfection button
, the Time Display and the Information
Display. When the program is complete, the machine is automatically
switched off.
Integrated Heat Disinfection
The AK 96 dialysis machine can be programmed to perform integrated
heat disinfection.
In this mode, the machine receives hot water from the central water
supply system when performing the heat disinfection program. The
heat disinfection program can be started manually, or preferably be
programmed to start automatically, during the heating of the central
water supply system.
The advantage with integrated heat disinfection, is that the tube
between the central water supply system (inlet water tube) and the
machine is also disinfected.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Hygiene and Maintenance
8:25
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...