Diascan PAUSED
during isolated UF
Appears:
During an attempt to activate the Diascan function
during on-going isolated UF phase or, when an attempt
to start a Diascan measurement check is made (single
or continuous measuring) during on-going isolated UF
phase.
Machine actions:
The Diascan measuring will automatically be paused
during the Isolated UF phase as the Diascan function
cannot perform measurement checks during the isolated
UF phase. Diascan measurement will start automatically
when the isolated UF phase is complete.
Measures:
-
Note that the Diascan measuring will automatically
be activated and rescheduled when the Isolated UF
phase is complete, without further attention alarms.
-
The attention alarm automatically disappears after
30 seconds.
Mute time:
Permanently muted buzzer.
Buzzer sound:
Soft buzzer.
14:42
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order