Recommended separation distances between portable and mobile RF communications equipment
and the AK 96 dialysis machine
The AK 96 dialysis machine is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the AK 96 dialysis machine can help
prevent electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the AK 96 dialysis machine as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
(m)
150kHz - 80MHz
80MHz - 800MHz
800MHz - 2500MHz
0,01
0.11
0.11
0.23
0,1
0.37
0.37
0.74
1
1.2
1.2
2.3
10
3.7
3.7
7.4
100
12
12
23
Rated maximum
output power of
mobile phone
-
-
2W
GSM/3G
-
-
0.33
For transmitters rated at a maximum output power not listed above, the recommended separation
distance
d
in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Note
•
At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
•
These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and
re
fl
ection from structures, objects and people.
Note
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Technical Data
9:21
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...