Attention Alarm List
AIR DETECTOR NOT ACTIVATED
Press AIR DETECTOR button
Appears:
During priming 15 seconds after the venous drip chamber
has been
fi
lled upon attention alarm request.
Machine actions:
The
Air detector button
will be
fl
ashing.
Measures:
Press the
fl
ashing
Air detector button
to activate the air
detector alarm function.
Mute time:
2 min.
Buzzer sound:
Soft buzzer
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:5
Attention alarms in alphabetical order