Acidic concentrate and liquid bicarbonate concentrate
1.
Attach Acidic concentrate
1.a.
Remove the red concentrate connector from the
stand-by port of the machine (compress the connector
before pulling the connector out).
1.b.
Attach the connector to a concentrate pick-up tube;
insert the connector into the pick-up tube port and push
it into place until it clicks in. Make sure that it clicks
in properly to prevent air leakage during treatment.
1.c.
Put the pick-up tube into the proper acidic concentrate
container (the blue concentrate connector must remain
in the stand-by port).
If SoftPac container for acidic concentrate is being used, connect
the red concentrate connector to the particular connector used for
SoftPac containers and insert it into the SoftPac container.
2.
Attach Liquid Bicarbonate concentrate
2.a.
Remove the blue concentrate connector from the
stand-by port of the machine (compress the connector
before pulling the connector out).
2.b.
Attach the connector to a concentrate pick-up tube;
insert the connector into the pick-up tube port and push
it into place until it clicks in. Make sure that it clicks
in properly to prevent air leakage during treatment.
2.c.
Put the pick-up tube into the proper liquid bicarbonate
concentrate container.
3.
Con
fi
rm concentrates
The con
fi
rm concentrate menus will automatically appear
when the red or blue concentrate connector is removed from
the stand-by port.
3.a.
Select
BICARBONATE
using the
Keypad
.
3.b.
Con
fi
rm that the connected concentrates correspond
with the displayed concentrate alternative on the
Information Display, using the
Select key
.
Note that at any time after con
fi
rming the concentrates, it
is possible to enter the concentrate menus to check/select
concentrates. Press the
Fluid path button
and select
CONC
. Go
into the sub menus and check/select the connected concentrates
using the
Keypad
.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Double Needle Treatment
4:11
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...