If a new treatment is to be started directly after the disinfection
program, press the
fl
ashing
Priming button
brie
fl
y and
PRESS SELECT 3 SECONDS TO START FCH
will be
displayed. Press the
Select key
for 3 seconds to start a new
function check and simultaneously reset previous values to zero.
Note especially that the chemical disinfectant connector cannot
be moved to the parking port during the chemical disinfectant
program.
Move the chemical disinfectant connector to the parking port
before next function check.
CAUTION
After the Chemical disinfection procedure a test for residues
must have been performed prior to connecting to a patient.
CAUTION
Filling the machine with chemical disinfectant
There may be a need to
fi
ll the machine
with a chemical disinfectant
during storage or when the AK 96 dialysis machine is planned to be
left unused for a long period of time. This function has to be preset
by an authorized technician, as well as the type of disinfectant. The
chemical disinfectant should preferably be peracetic acid.
In order to perform a
fi
lling of the machine with chemical disinfectant
follow the instructions in "Chemical Disinfection Program -
Performing" on page 8:27 previously in this section from the beginning
up to point 7. Continue as follows:
7. Press the
On/Off button
during the chemical disinfectant
fi
lling phase. The machine will continue to run the program
for some minutes after the
On/Off button
has been pressed.
This is to
fi
ll up all the tubes and to even out the concentration
of the disinfectant. It will then be automatically switched off.
Now the machine is
fi
lled with disinfectant.
Note especially that the disinfectant connector must remain in
the disinfectant container/disinfectant port and not be moved
to the parking port before the chemical disinfectant program is
complete. This is due that the chemical disinfectant is pushed
back to the disinfectant container to empty the connector/intake
tube.
8.
When the machine is to be used again
and the
On/Off button
is pressed, the machine will start a
rinse/disinfection program as soon as it is switched on. The
Rinse/Disinfection button
will be lit at this moment. The
remaining time will be shown on the time display.
9. When the program is complete and the machine has been
drained, the
Priming button
starts to
fl
ash.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Hygiene and Maintenance
8:29
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...