Profiling of ultrafiltration rate
For ultra
fi
ltration rate, 3 different graphs can be set. Linear increasing
or decreasing graph, increasing or decreasing graph - in steps, or graph
where the UF rate changes at intervals. All graphs represent changes of
the UF rate throughout the treatment.
The change of UF rate for the linear graph is determined by the
following settings; total UF volume, treatment time and start UF rate.
Based on these parameters, the machine will perform a smooth and
constant adjustment of the UF rate during the treatment time when
pro
fi
ling has been activated. See the following
fi
gure.
Figure 7:2
Decreasing linear graph for UF rate pro
fi
ling
The change of UF rate for the graph in steps is determined by the
following settings; total UF volume, treatment time, number of steps in
the graph and start UF rate. Based on these parameters, the machine
will perform adjustments of the UF rate in steps during the treatment
time when pro
fi
ling has been activated. See the following
fi
gure.
Figure 7:3
Decreasing graph in steps for UF rate pro
fi
ling
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Profiling
7:3
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...