TREATMENT WILL STOP IN 2 MINUTES
HIGH VENOUS PRESS TEST NOT MADE
Appears:
After 8 minutes in treatment, if the machine has not been
able to test the venous pressure measurement function.
This is due to that the venous pressure has not exceeded
50 mmHg during the priming procedure.
Machine actions:
The blood pump and the ultra
fi
ltration will be
automatically stopped 2 minutes after this attention alarm
has been generated.
Measures:
Discontinue treatment. Contact authorized technician.
Mute time:
2 min.
Buzzer sound:
Continuous buzzer
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:121
Attention alarms in alphabetical order