8.e.
Make sure that chosen syringe type matches displayed
syringe type
in the
HEPARIN MENU
(press the
Blood Path button
, select
HEPARIN
) If not call
authorized service technician.
8.f.
Insert the syringe in the pump; the plastic collar at the
end of the syringe must
fi
t in the groove of the pump,
in the corresponding
fi
gure.
8.g.
Push back the piston and insert the plastic plate on
the plunger in the groove by the locking wheel,
in
the
fi
gure.
8.h.
Turn the locking wheel upwards until resistance is felt.
This is to lock the plastic plate of the plunger in the
groove,
in the
fi
gure.
8.i.
Check that the syringe is properly secured by lightly
pulling on the plunger,
in the
fi
gure.
8.j.
Press the
Blood Path button
, the
BLOOD PATH MENU
will be displayed on
the Information Display.
8.k.
Select
HEPARIN
using the
Keypad
, the
HEPARIN MENU
will be displayed.
8.l.
If a bolus dose of heparin solution is to be administered
at treatment start, set a bolus volume in ml (
BOLUS
)
suitable for the patient using the
Keypad
.
8.m. Set a heparin solution
fl
ow rate in ml/hour
(
FLOW RATE
) suitable for the patient. Note that
during priming it is not possible to run the heparin
pump with normal
FLOW RATE
.
4:26
AK 96
®
Operator's Manual - Double Needle Treatment
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...