DISCONNECT VENOUS PRESSURE
Press VENOUS button to confirm
Appears:
When the venous pressure transducer of the venous blood
line has been attached too early in function check. See
"Venous Blood Line - Attach" on page 4:28 in part 1 for
information.
Machine actions:
None.
Measures:
Disconnect the venous pressure transducer and press
Venous button
to con
fi
rm.
If the attention alarm is still generated, consult an
authorized technician.
Mute time:
5 min.
Buzzer sound:
Soft buzzer