FLUID TUBES NOT IN SAFETY BYPASS
Connect tubes to safety bypass
Appears:
When the dialysis
fl
uid tubes have been removed from
the safety couplings, during function check or during
disinfection program.
Machine actions:
Function check is interrupted.
Measures:
Reconnect the dialysis
fl
uid tubes to the safety couplings.
If a disinfection or rinse program has been started, check
that the dialysis
fl
uid tubes are properly connected to the
safety couplings.
Mute time:
2 min.
Buzzer sound:
Soft buzzer
14:62
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order