Appears:
The transmembrane pressure (TMP) has passed the set
TMP low alarm limit.
Machine actions:
None.
Possible cause:
The TMP value during treatment depends mainly on the
following parameters; the set treatment time, the set UF
volume and the set blood
fl
ow rate. Changes of these
three parameters affect the TMP value, as well as changed
performance of the dialyzer as the treatment proceeds.
Changes in the machine UF control system can also cause
reduced accuracy in measuring the patient's weight loss.
The following points may cause low TMP alarm:
-
changes in the UF-control system of the machine
-
the set treatment time has been increased
-
the set UF volume value has been decreased
-
increased blood
fl
ow rate
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Alarms