Manual Priming and Rinsing Procedure
Check Before Priming
Check that the priming
fl
uid and waste bag are properly connected.
Remove any obstructing clamps.
Check that the venous blood line is correctly placed in the priming
detector. See "Venous Blood Line - Attach", point 5, in previous
instructions.
Check that the pressure transducers of the blood lines have been
properly attached to the pressure connectors of the machine.
Check all connections.
Check that the priming UF rate is correctly set; press
the
Fluid Path button
, select
UF
and check/adjust
SET PRIMING UFR
using the
Keypad
. Follow the instruction
for use of the dialyzer.
Note
•
The
CONCENTRATE STANDBY MODE
menu, displayed on
the Information Display, indicates that concentrate stand-by
mode is active. Resume dialysis
fl
uid preparation (deactivate
concentrate stand-by mode) by pressing the
Select key
for three
seconds, as the menu requests. The dialysis
fl
uid will be ready
within approximately 2 minutes.
•
Always follow the manufacturer's recommendations for the
minimum rinsing volume (for the blood lines and the dialyzer).
Note
Priming Procedure
This menu will be displayed on the Information
Display when manual priming is active. If the
ASSISTED PRIMING SETUP MENU
is displayed,
brie
fl
y press the lit
Priming button
and select
PRIMING
using
the
Keypad
. Then press the
Select key
for three seconds to
activate manual priming mode.
1. Check that the
Priming button
is lit.
Follow the manufacturer's instructions for use of the dialyzer
and the blood lines.
2. Press the
fl
ashing
Blood pump button
. The pump speed
(
PUMP SPEED
) will automatically start at 100
ml/min.
Adjust as necessary using the
Blood Pump Up/Down keys
;
the blood
fl
ow menu will be displayed, adjust
SET QB
. If
the blood pump segment has been automatically threaded,
check for possible kinks.
4:40
AK 96
®
Operator's Manual - Double Needle Treatment
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...