Diascan
®
Clearance measurement is only available if the optional Diascan is
installed. During UF STEP and INTERVAL pro
fi
ling, Diascan system
is not available. The speci
fi
cation is based on in vitro measurements
with saline.
Clearance K typical
precision
± 8% (± 1SD). Precision has been
validated in HD double needle, for blood
fl
ows 200 to 400 ml/min and
fl
uid
fl
ows
500 to 700 ml/min.
Cumulated water volume
cleared of Kt
0 to 100 l (± 6% (± 1SD, based on 7
measurements)
Dialysis dose Kt/V
0 to 3
Measurement interval
15, 30 or 60 minutes
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Technical Data
9:7