Performance and specification -
Control System
Blood Flow Control
Values for the blood pump(s) are based on a pressure of -150 mmHg
before the arterial blood pump with a pump segment of 7.9 mm and
2.0 mm wall thickness.
For 6.35 mm pump segment it is possible to set the blood
fl
ow to
15 ml/min.
For pediatric blood tubes with pump segment of 4.0 mm it is possible
to set blood
fl
ow to 5, 10 or 15 ml/min. Accuracy is then ±5 ml/min.
Double Needle
Flow rate
0 and 20 to 500 ml/min (-150 mmHg
pre pump pressure, 0 to 500 mmHg post
pump pressure, pump segment diameter
7.9 mm)
Flow accuracy
±10 ml/min or ±10 %, whichever is
largest
Accumulated blood volume 0 - 327 liters
Volume accuracy
±0.6 l * treatment time (h) or ±10 %
Single Needle
Single needle is only available if the optional arterial clamp is installed.
Arterial
fl
ow rate
0 and 20 to 500 ml/min (-150 mmHg
pre pump pressure, 0 to 500 mmHg post
pump pressure, pump segment diameter
7.9 mm)
Flow accuracy
±10 ml/min or ±10 %, whichever is
largest
Pressure control
10 to 500 mmHg (±50 mmHg), venous
pressure control
Accumulated blood volume 0 - 327 liters
Volume accuracy
±0.6 l * treatment time (h) or ±10 %
9:2
AK 96
®
Operator's Manual - Technical Data
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...