If a new calculation is desired, select
WATSONFORMULA
and it will be possible to
redo the procedure from the start.
Alarm limit for Kt/V
To set the alarm limit for minimum Kt/V, select
Kt/V LIMIT
and set a proper alarm limit value
using
the
Keypad
.
For further information; see "Kt/V alarm limit", page
Target for Kt/V
To set a desired dialysis dose target value (Kt/V)
which is to be achieved at treatment end, select
Kt/V TARGET
and set a proper value using the keypad.
The target value for Kt/V can be set between
OFF
and 2.0.
When the Kt/V target value has been set, the machine
calculates the time needed to reach the target. This time
to target (
TTT
) will be displayed in the viewing menu as
soon as the
fi
rst Diascan measurement check has been
performed.
12:16
AK 96
®
Operator's Manual - Diascan
HCEN12239 Revision 12.2010
Program version 3.xx