FCH PROLONGED DUE TO CHEM TESTS
Do not connect conc. or BiCart
Appears:
In function check if a chemical disinfection program has
previously been performed.
Machine actions:
Ongoing chemical valve leakage test.
Measures:
Wait for the machine test to be
fi
nished, to be able to
connect the concentrates. It is possible when the attention
alarm text has disappeared.
Mute time:
-
Buzzer sound:
No buzzer
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:53
Attention alarms in alphabetical order