Transmembrane Pressure
From the Fluid Path main menu; select
TMP
by using the
Keypad
to
reach the sub menu.
The following will be shown on the Information Display:
SET LOW
; low transmembrane pressure alarm limit
TMP
; current transmembrane pressure
SET HIGH
; high transmembrane pressure alarm limit
In this display, upper cursors will be placed on both the low and the
high transmembrane pressure alarm limits.
To move the complete alarm window
(both low and high alarm limits
will change simultaneously),
fi
rst press the
Select key
. Then move
the window on the scale using the
Up
and
Down Display keys
. Close
using the
Select key
.
To adjust one alarm limit at a time
,
fi
rst press the
Left
or
Right Cursor key
to select alarm limit. Then press the
Select key
to open
the position. Change the value using the
Up
and
Down Display keys
.
Close using the
Select key
.
3:18
AK 96
®
Operator's Manual - Operating the Machine
HCEN12239 Revision 12.2010
Program version 3.xx