7.
Arterial Pressure Connector
Attach the arterial pressure transducer protector of the
blood line to the arterial pressure connector; press and
simultaneously turn.
Check all necessary clamps
If the heparin pump is not to be used, especially ensure that
the cap at the end of the thin line used for heparin solution, is
properly closed.
8.
Heparin Pump
The heparin pump is used for continuous and/or bolus
administration of heparin. Heparin is to be mixed (diluted)
with the proper solution in a correct syringe used for the
purpose. The Heparin pump can be programmed for different
syringe sizes , syringes must comply with ISO 7886-2.
Consult the authorized technician for information concerning
heparin pump presets for the particular machine. Make sure
that chosen syringe type matches displayed syringe type
in the
HEPARIN MENU
, if not call authorized service
technician.
See the following pages (continuing point 8) for instructions on
how to attach the syringe to the heparin pump and how to set the
heparin pump values in the machine.
4:24
AK 96
®
Operator's Manual - Double Needle Treatment
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...