Distribution volume
The operator is requested to set distribution volume in
litres, estimated for the individual patient, in order to
obtain a correct
Kt/V
value from the Diascan function.
The distribution volume has to be properly estimated
and set by the operator. The calculation of Kt/V done
in the machine is entirely dependent on the value for
the individual patient being the correct one. For further
information; see "Glossary of Diascan® parameters and
key terms", page 12:4.
If desired, it is possible for the machine to carry out
a distribution volume calculation using the Watson
formula, see further on in this section for instructions.
Distribution volume setting
To set a distribution volume, select
VOLUME
and set a
proper value
using the
Keypad
. 40 l is the default
starting point value ; adjust the value as necessary.
Distribution volume calculation and setting using the
Watson formula
It is possible for the machine to carry out a distribution
volume calculation, displayed in litres, using the Watson
formula. In order for the machine to provide a value, the
gender, weight, height and age (if male) of the particular
patient must be entered into the machine. When the
calculation has been done, a (W) will be displayed beside
the distribution volume value provided, showing that the
value has been calculated by the machine.
12:14
AK 96
®
Operator's Manual - Diascan
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...