4. Disconnect the inlet dialysis
fl
uid tube from the safety
coupling of the machine; remove it in the same way as the
outlet dialysis
fl
uid tube described in the former point 1
in these instructions. The spent dialysis
fl
uid
fl
ows
to the
machine from the dialyzer
via this tube. A small arrow,
fi
xed just below where the tube comes out from the machine,
shows the
fl
ow direction.
5. Keep the button on the dialysis
fl
uid tube pressed and hold
while attaching it to the dialyzer connector as shown in the
corresponding
fi
gure.
6. Release the button and simultaneously push the connector
into place until it clicks in. Make sure that it clicks in
properly to prevent leakage during treatment.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Double Needle Treatment
4:37