INCORRECT ACIDIC CONCENTRATE
Check concentrate
Appears:
When the selected concentrate does not match the
concentrate connected to the machine.
Machine actions:
The dialysis
fl
uid is bypassed from the dialyzer.
Measures:
Check concentrate selection and that the proper
concentrate has been connected. See "Connect/Con
Concentrates" on page 4:7 in part 1 for instructions.
Mute time:
2 min.
Buzzer sound:
Soft buzzer
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:81
Attention alarms in alphabetical order