Single Needle Stroke Volume Alarm
Appears:
In single needle treatment when the measured stroke
volume
STROKE
is lower than the set stroke volume
SET
.
Machine actions:
None.
Possible cause:
-
Decreased performance of the patient's blood access.
-
Adjustments to optimize the single needle treatment
may have been overlooked.