Appears:
The dialysis
fl
uid temperature has passed the set
temperature low alarm limit.
Machine actions:
-
The dialysis
fl
uid is bypassed from the dialyzer.
When the dialysis
fl
uid temperature is within the set
alarm limits again, the dialysis
fl
uid automatically enters
the dialyzer.
Possible cause:
-
The machine has not yet stabilized the dialysis
fl
uid
temperature after an interruption of the water supply
to the machine.
-
The power supply to the heating rods in the machine
has been interrupted.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Alarms