After Treatment
Follow the instructions in chapter 4, "Start the Treatment” in part
1 to start the treatment. At any time during the ongoing treatment,
do the settings for the isolated UF phase according to the following
instructions.
Bear in mind that the time and UF volume in isolated UF are
automatically added to the diffusion phase and consequently increase
the complete treatment time and the total UF volume.
Procedure
Use the
Keypad
next to the Information Display to select options and
to navigate within and between the menus. For details concerning the
keys included in the
Key pad
, see "Keypad" in chapter 3. part 1
1. Press the
Fluid Path button
, select
UF
.
2. Select
ISOL UF
and enter the sub menu.
3. Set the ultra
fi
ltration volume,
SET ISOL UF
, for the
isolated UF phase.
4. Set the time,
SET TIME
, for the isolated UF phase.
5. Press the
Back key
for 3 seconds and the Treatment Overview
Menu will resume. The Isolated UF settings will be kept.
When the diffusion phase is complete an attention alarm will
appear:
6. Check that the previously set values are the correct ones and
activate isolated UF by pressing the
Select Key
for 3 seconds.
Isolated UF will now be activated.
Note that if the heparin pump is used, it will run during the
complete treatment irrespective of the added isolated UF
phase and keep the settings for
BOLUS
,
FLOW RATE
and
STOP LIMIT
. This means that the stop time will be adjusted in
accordance with the time for the added isolated UF phase.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Isolated UF
6:5
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...