Dialysis fluid preparation
Pressure regulators
After pressure regulator
PR1 and the heat
exchangers
80 kPa (0.8 ±0.1 bar)
After pressure regulator
PR2
130 ±10 mmHg
Temperature
Temperature
Adjustable 33 to 40 °C.
Accuracy is valid only if dialysis
fl
uid
temperature is greater or equal to ambient
temperature.
Alarm limits
Adjustable 33 to 40 °C
Accuracy
+0.5/-1.5 °C (+1.0/-2.5 °C with UFD) at
the dialysis
fl
uid outlet from the machine.
Heater capacity
1300 W (+10 % / -5 %) at 115 V
3 X 580 W (+10 % / -5 %) at 230 V
Overheat protection
Reg-temp 80 °C or CondA-temp 70 °C in
treatment, software
Reg-temp 99 °C in disinfection, software
The heater can only be on if there is a
fl
ow through the heater. The
fl
ow is
detected by a
fl
ow switch.
Flow rate
Dialysis Fluid Flow Rate
300 to 700 ml/min in steps of 20 ml/min
Accuracy
±10 % or 50 ml/min whichever is largest
Degassing
By use of negative pressure, -610 mmHg.
Adjustable degassing pressure between -300 and -700 mmHg
(-650 mmHg with 700 ml/min
fl
ow rate).
Accuracy
±40 mmHg
Dialysis
fl
uid pressure
Dialysis
fl
uid pressure
-400 to +300 mmHg
Accuracy
±10 mmHg or ±5 %
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Technical Data
9:5
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...