NEGATIVE UF RATE
Check cause and press SELECT key
Appears:
When there is a back
fi
ltration of 500 ml to the patient or
a back
fi
ltration rate >50 ml/min for 5 min.
Machine actions:
None.
Measures:
Check the outlet tube (the tube where the spent dialysis
fl
uid
fl
ows out from the machine) for obstructions.
If there are obstructions, remove them and press the
Select key
to con
fi
rm the removal. Check the UF rate
is correct before leaving the machine, if not call an
authorized technician.
Mute time:
2 min.
Buzzer sound:
Soft buzzer
CAUTION
Incorrect weight loss may occur if the machine is left
for a longer period with this attention.
CAUTION
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:93
Attention alarms in alphabetical order