NEW ULTRA FILTER REQUIRED
days since last change.
###
Appears:
When the ultra
fi
lter needs to be replaced because the
limit
for number of days since last change has been
exceeded.
Machine actions:
None.
Measures:
Replace the ultra
fi
lter and con
fi
rm that the ultra
fi
lter has
been replaced.
Mute time:
Permanently muted buzzer.
Buzzer sound:
Supersoft buzzer.
14:94
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order