Diascan FAILURE
Turn OFF Diascan
Appears:
Appears when too many Diascan steps have been done
during that particular treatment.
Machine actions:
None.
Measures:
Performing further Diascan steps may lead to an
imbalance in the sodium level of the patient. The
Diascan function is not permitted to perform any more
measurement checks. Turn Diascan function off. Contact
an authorized technician.
If this attention alarm appears frequently, discuss preset
limits with an authorized technician.
Mute time:
Permanently muted buzzer.
Buzzer sound:
Soft buzzer
14:38
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order