Automatic start of the Low Flow Heat phase
1. Press the
Rinse/Disinfection button
brie
fl
y.
2. The
RINSE/DISINFECTION MENU
will be displayed on
the Information Display.
3. Select
HEAT
.
4. Check that
HEAT
is displayed as selected program. If
HEAT
is displayed, continue to point 4. If not, do point 4.a
and 4.b before continuing to point 5.
4.a.
Select
CHANGE
.
4.b.
Select
HEAT
.
5. Start the program by pressing the
Select key
for 3 seconds
until the
Rinse/Disinfection button
lights up.
6. Respond to the text shown on the Information Display on
whether to include the WRO or not. Use the alternative to
include the WRO.
7. The process will continue automatically from here. When
the heat disinfection of the AK 96 dialysis machine and the
low
fl
ow heat are
fi
nished the
Priming button
will start to
fl
ash (not if an automatic switch off has been done). The
WRO will perform a heat disinfection of its
fl
uid path after
the low
fl
ow heat.
Note that if a new function check is started during the WRO heat
process an attention is generated during the function check due to
8:20
AK 96
®
Operator's Manual - Hygiene and Maintenance
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...