NO CUFF ATTACHED
Attach or check BPM cuff
Appears:
When the machine registers that no cuff is applied to the
patient.
Machine actions:
Buzzer delayed 2 min.
Measures:
Check for kinks or leakage from the cuff and the cuff
hose. Check also that the cuff hose is properly connected
to the machine.
Mute time:
Permanently muted buzzer.
Buzzer sound:
Soft buzzer
14:100
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order