DIASCAN IS NOT RESTARTED
Check settings and activate
Appears:
When treatment is resumed after the dialyzer and the
blood lines have been changed during treatment.
Machine actions:
The Diascan function has been automatically deactivated.
Measures:
Press the
Fluid Path button
, select
DIASCAN
and then
SET DIASCAN
using the
Keypad
. Check that the set
Diascan parameter values are the correct ones, adjust if
necessary. Restart the Diascan function by pressing the
Select key
for 3 seconds.
Mute time:
2 min.
Buzzer sound:
Soft buzzer
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:39
Attention alarms in alphabetical order