BLOOD PUMP OVERLOAD IN FCH
Check blood pump, Press SELECT
Appears:
When it is hardly possible for the blood pump to run.
Machine actions:
The blood pump stops.
Measures:
Check that the blood pump segment is correctly
positioned, see Arterial Blood Line - Attach on page 4:16
in part 1 for instructions. When properly positioned,
press the
Select key
.
If the attention alarm appears again, call an authorized
technician.
Mute time:
Permanently muted buzzer.
Buzzer sound:
Soft buzzer
14:22
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order