Isolated UF - Performing
Before Treatment
Follow the instructions in chapter 4, "Start the Treatment" in part 1.
When the settings in "Set Treatment Parameters" on page 4:52 in part
1 are being made, set the time and UF volume for the diffusion phase
only. Thereafter, make the settings for the isolated UF phase according
to the following instructions.
Bear in mind that the time and UF volume in isolated UF are
automatically added to the diffusion phase and consequently increase
the complete treatment time and the total UF volume.
Procedure
Use the
Keypad
next to the Information Display to select options and
to navigate within and between the menus. For details concerning the
keys included in the
Key pad
, see "Keypad" in chapter 3. part 1
1. Press the
Fluid Path button
, select
UF
.
2. Select
ISOL UF
and enter the sub menu.
3. Set the ultra
fi
ltration volume,
SET ISOL UF
, for the
isolated UF phase.
4. Set the time,
SET TIME
, for the isolated UF phase.
5. Activate isolated UF by pressing the
Select Key
for 3
seconds. Isolated UF will automatically be activated when
the
UF Start/Stop button
is being pressed at treatment start.
Start the treatment according to the instructions in "Connect the
Patient" on page 4:53 and "Start the Treatment" on page 4:56 in
part 1.
Note that if the heparin pump is used, it will run during the
complete treatment (both diffusion and isolated UF phase) and
keep the settings for
BOLUS
,
FLOW RATE
and
STOP LIMIT
.
Activated isolated UF is displayed at the activity
fi
eld of the
Treatment Overview Menu on the Information display. During
the phase, the accumulated values for ultra
fi
ltration
ACC UF
and
time
ACC TIME
will be displayed on the isolated UF menu.
Press the
Fluid Path button
, select
UF
and then
ISOL UF
to
view the menu.
If necessary, isolated UF can be deactivated at any time during the
phase, by pressing the
Select Key
for 3 seconds when the isolated UF
menu is displayed.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Isolated UF
6:3
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...