Venous Pressure Alarms
Appears:
The venous pressure has reached the set venous pressure
high alarm limit.
Machine actions:
-
The blood pump stops.
-
The arterial blood line clamp closes in single needle
mode.
-
The ultra
fi
ltration rate is set to zero.
When the venous pressure is within the set alarm limits
again, the blood pump automatically starts.
Possible cause:
-
The venous blood line is kinked or clamped.
-
The position of the venous needle has changed.
-
Clotting has occurred in the venous blood line after
the venous pressure measurement point; in the blood
line, in the venous drip chamber, or in the needle.
13:14
AK 96
®
Operator's Manual - Alarms
HCEN12239 Revision 12.2010
Program version 3.xx