FCH IS WAITING
Attach conc connectors to ports
Appears:
During function check if a chemical disinfection program
has previously been performed and the concentrate
connectors (or disinfectant connector) are not properly
connected to the machine.
Machine actions:
Function check is interrupted.
Measures:
Check that the concentrate connectors are properly
connected to the stand-by ports. Check that the
disinfectant connector is properly connected to the
parking port. It is not possible to connect the concentrates
yet.
Mute time:
2 min.
Buzzer sound:
Soft buzzer
14:50