DIALYSIS FLUID NOT READY FOR
TREATMENT. FLUID READY IN $## s
Appears:
When the dialysis
fl
uid tubes are disconnected from
the machine too soon after the function check has been
fi
nished.
Machine actions:
Automatic
fl
ush of the
fl
ow path where the ultra
fi
lter is
placed is performed.
Measures:
Wait until the attention alarm has disappeared and the
bypass path of the Flow Diagram lights up green before
disconnecting the dialysis
fl
uid tubes, as there is a high
fl
uid
fl
ow through the
fl
uid tubes. The period of time for
the appearance of the attention alarm is dependent on the
dialysis
fl
uid
fl
ow rate
.
Mute time:
-
Buzzer sound:
No buzzer
14:36
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order