Details
1.
Blood Pump
Open the cover of the blood pump by pressing the middle
of the cover at the same time as the cover is being opened.
After this, the blood pump segment can be automatically
or manually threaded.
If the blood pump is automatically threaded, the machine
will thread the blood pump segment over the blood pump
rotor either during function check or when priming is started
dependent on when the pump segment has been placed over
the blood pump rotor.
If it is manually threaded, the operator threads the segment
through the blood pump during the arterial blood line set-up.
Instructions on both alternatives follow after this.
If automatically threaded
Place the blood pump segment over the blood pump rotor
as shown in the corresponding
fi
gure.
While keeping the segment in place, close the blood pump
cover by pressing at the point shown in the corresponding
fi
gure.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Double Needle Treatment